AP100 LN2 Ultrapure Fogger
$0.00
AP100 LN2 Ultrapure Fogger with rolling, stainless steel enclosure, ≈ 16 M³ / Min. ultrapure fog, ≈ 21 ft. visual distance, ≈ 1280 M³ fog during typical 80 minute operation cycle to visualize airflow turbulence in large Clean Rooms, ISO Suites and Barrier Isolators. Two 80mm fog outlets with velocity control and volume control.
Smoke Study, Smoke Test or Airflow Visualization Study | AP100 LN2 Ultrapure Fogger (Part # AP100-4) |
---|---|
Fogger Design | AP100 has a dense, Ultrapure fog with 2-3 micron fog droplets using LN2+DI Water, or WFI or Sterile Water. 316L stainless-steel enclosure with Two 80mm Fog Outlets, 13.5 Liters water and 100 L LN2 |
Fog Application | Fog in medium sized pharmaceutical clean rooms, sterile rooms, ISO suites. Also use in Stage rooms and clean rooms. Two 80mm Fog Outlet with 80mm hose options |
Fog Volume | 16 Cubic Meters ultrapure fog/minute |
Fog Duration | about 80 minutes at max fog volume; or 120 minutes at lowest fog volume |
Visual Airflow Distance | 21 feet using the dual 80mm fog outlets |
Liquids Used | 100 L LN2 + 13.5 L DI Water, Sterile Water or WFI Water |
Optional Accessories | 1.2 M x 80mm Fog Wand with 90 degree outlet nozzles, Wireless Remote Control, 5M x 80mm Fog Hose, 5M x 50mm fog hose, Accessory Rolling Carry Case, 0.6M x 50mm Fog Rake, 80mm Y Adaptor with Butterfly Valves, LED Fog Contrast Light, 350mm Fog Nozzle |
Power | 100-240V 47/63HZ. Power Cords for EU, US, UK, CH, SG, PRC, JP, AUS |
AP100 LN2 Ultrapure Fogger (Part # AP100-4)
AP100 LN2 Ultrapure Fogger is used in smoke studies and airflow visualization Studies in medium sized semiconductor clean rooms, medium sized pharmaceutical ISO Suites and industrial clean rooms. Ultrapure foggers combine nitrogen and water vapor to form a dense, ultrapure fog to visualize airflow turbulence; using liquid nitrogen and water (De-ionized water, WFI [water for injection] or sterile water. Below is a selection of foggers and performance details for each fogger, which you can match to your requirements for your smoke studies.
Fogger Design | AP100 LN2 Ultrapure Fogger has a dense, Ultrapure fog with 2-3 micron fog droplets using LN2+DI Water, or WFI or Sterile Water. 316L stainless-steel enclosure with One 80mm Fog Outlet, 13.5 Liters water and 100 Liters LN2 |
Fog Application | Fog in pharmaceutical clean rooms, sterile rooms, ISO suites. Also use in Stage rooms and clean rooms. Two 80mm Fog Outlets with 80mm and 50mm fog hose options |
Fog Volume | 16 Cubic Meters ultrapure fog/minute |
Fog Duration | about 80 minutes at max fog volume; or about 120 minutes at lowest fog volume |
Visual Airflow Distance | about 21 feet using the two 80mm fog outlets |
Liquids Used | LN2+DI Water, Sterile Water or WFI Water |
Optional Accessories | 1.2 M x 80mm Fog Wand with 90 Degree outlet nozzles, Wireless Remote Control, 5M x 80mm Fog Hose, Accessory Rolling Carry Case, 5M x 80mm fog Hose, 80mm Y Adaptor with Butterfly Valves, LED Fog Contrast Light, 350mm Fog Nozzle |
Power | 100-240V 47/63HZ. Power Cords for EU, US, UK, CH, SG, PRC, JP, AUS |
LN2 Ultrapure Fogger such as the AP100 use 13.5 Liters of Di-Water, Sterile Water, or WFI Water and 100 Liter LN2. The Fog is produced by allowing the LN2 to boil, which occurs at room temperatures, thus releasing a nitrogen vapor. The nitrogen vapor is directed into the water cavity, which has a dual water heater controlling water temperature from 55C to 70C. The heated water emits water vapor, which combines with the nitrogen vapor to form a very dense, highly visible ultrapure fog. This fog is directed into your airflow for airflow visualization studies, commonly known as smoke studies. The ultrapure fog is composed of 2-3 micron fog vapor droplets, and is very dense in visibility. Ultrapure fog is used to visualize airflow turbulence, airflow patterns, airflow uniformity and to locate dead airflow zones in clean rooms and around equipment. Compare the different LN2 foggers for performance below.
Smoke Studies, Cleanroom Fogger Comparisons | ||||||||||
Fogger Model Number | Enclosure Material | Fog Application | Fog Volume Minutes | Duration in Minutes | Visible Distance | Appx. Fog Volume | Fog Outlets | Velocity Control | Volume Control | Wireless Control |
SMST-12* | Polyethylene | Airflow & Fume Hoods | ≈ 0.5 M³ | ≈ 45 Min., Pure Fog | ≈ 7 Ft. | ≈ 22.5 M³ | One 80mm | Yes | No | Optional |
AFM Cube* | 316 Stainless | Gloveboxes, BSC | ≈ 0.7 M³ | ≈ 45 Min., Pure Fog | ≈ 8 Ft. | ≈ 32 M³ | One 80 | Yes | Yes | No |
AFM24 (CRF3)* | 316 Stainless | Gloveboxes, BSC | ≈ 0.83 M³ | ≈ 55 Min., Pure Fog | ≈ 9 Ft. | ≈ 45 M³ | One 80 | Yes | Yes | Optional |
AFM30* | 316 Stainless | BSC & Barrier Isolator | ≈ 1 M³ | ≈ 45Min., Pure Fog | ≈ 10 Ft. | ≈ 45 M³ | One 80 | Yes | Yes | Optional |
AFM35 (CRF4)* | 316 Stainless | BSC & Barrier Isolator | ≈ 1.4 M³ | ≈ 45 Min., Pure Fog | ≈ 11 Ft. | ≈ 63 M³ | One 80 | Yes | Yes | Optional |
AFM40 (CRF6) * | 316 Stainless | Barrier Iso-lator, ISO Suite | ≈ 2.0 M³ | ≈ 35 Min., Pure Fog | ≈ 13 Ft. | ≈ 70 M³ | Two 80s | Yes | Yes | Optional |
* Cleanroom Fogger in Smoke Studies uses De-Ionized Water, Water for Injection or Sterile Water | ||||||||||
Fogger Model Number | Enclosure Material | Fog Application | Fog Volume Minute | Duration in Minutes | Visible Distance | Appx. Fog Volume | Fog Outlets | Velocity Control | Volume Control | Wireless Control |
M2001, M2100 LN2 ** | 316 Stainless | BSC & Barrier Isolator | ≈ 1.5 M³ | ≈ 35 Min., Ultrapure Fog | ≈ 10 Ft. | ≈ 52 M³ | One 60mm | No | No | No |
AP35 LN2** | 316 Stainless | Small Cleanroom, ISO Suite | ≈ 5 M³ | ≈ 45 Min., Ultrapure Fog | ≈ 15 Ft. | ≈ 225 M³ | One 80mm | Yes | Yes | Optional |
AP50 LN2** | 316 Stainless | Small Cleanroom, ISO Suite | ≈ 8 M³ | ≈ 60 Min., Ultrapure Fog | ≈ 18 Ft. | ≈ 480 M³ | One 80mm | Yes | Yes | Optional |
AP100 LN2** | 316 Stainless | Medium Cleanroom, ISO Suite | ≈ 16 M³ | ≈ 80 Min., Ultrapure Fog | ≈ 21 Ft. | ≈ 1280 M³ | Two 80s | Yes | Yes | Optional |
AP150 LN2** | 316 Stainless | Large Cleanroom, ISO Suite | ≈ 24 M³ | ≈ 120 Min Ultrapure Fog | ≈ 25 Ft. | ≈ 2880 M³ | Three 80s | Yes | Yes | Optional |
** Ultrapure LN2 Cleanroom Fogger in Smoke Studies uses LN2 + De-Ionized Water, Water for Injection or Sterile Water |
Additional information
Smoke Study, Smoke Test or Airflow Visualization Study | AP100 LN2 Ultrapure Fogger (Part # AP100-4) |
---|---|
Fogger Design | AP100 has a dense, Ultrapure fog with 2-3 micron fog droplets using LN2+DI Water, or WFI or Sterile Water. 316L stainless-steel enclosure with Two 80mm Fog Outlets, 13.5 Liters water and 100 L LN2 |
Fog Application | Fog in medium sized pharmaceutical clean rooms, sterile rooms, ISO suites. Also use in Stage rooms and clean rooms. Two 80mm Fog Outlet with 80mm hose options |
Fog Volume | 16 Cubic Meters ultrapure fog/minute |
Fog Duration | about 80 minutes at max fog volume; or 120 minutes at lowest fog volume |
Visual Airflow Distance | 21 feet using the dual 80mm fog outlets |
Liquids Used | 100 L LN2 + 13.5 L DI Water, Sterile Water or WFI Water |
Optional Accessories | 1.2 M x 80mm Fog Wand with 90 degree outlet nozzles, Wireless Remote Control, 5M x 80mm Fog Hose, 5M x 50mm fog hose, Accessory Rolling Carry Case, 0.6M x 50mm Fog Rake, 80mm Y Adaptor with Butterfly Valves, LED Fog Contrast Light, 350mm Fog Nozzle |
Power | 100-240V 47/63HZ. Power Cords for EU, US, UK, CH, SG, PRC, JP, AUS |